Features: · Lightweight, hand-held design· Easy-to-read LCD display· Simple operation· High accuracy peak reading circuit· Wide 10-1000 V test range· Precise auto-sensing shutoffApplication· Gas tubes· Carbon gaps· MOVs· Zener and avalanche diodes· Thyristor devices· Finished productDe...
4010-01: Features: · Lightweight, hand-held design· Easy-to-read LCD display· Simple operation· High accuracy peak reading circuit· Wide 10-1000 V test range· Precise auto-sensing shutoffApplication· ...
SeekIC Buyer Protection PLUS - newly updated for 2013!
268 Transactions
All payment methods are secure and covered by SeekIC Buyer Protection PLUS.
Bourns® 4010-01 Surge Protector Test Set is a hand-held, battery-operated tester designed to measure the clamping voltage and dc breakdown voltage of most surge protective devices. The 4010-01 is suitable for testing gas tubes, carbon gaps, MOVs, Zener and avalanche diodes and thyristor devices, both as components and as complete protectors.
The 4010-01 is a true slow-rise tester because it does not use a pulse transformer. It contains unique test circuitry and a precise 1mA detector. As soon as 1mA is conducted through clamping type devices, such as MOVs and avalanche diodes, the control circuits removes the test voltage from the output terminals and displays the clamping voltage on the meter. A low energy capacitive circuit will discharge through crowbar type protectors, such as gas tubes and carbon blocks, and trigger the detector. The breakdown voltage of the 4010-01 will be displayed and the test voltage removed.